Radon Detectors for New York State Residents

Active 2140608 RFPLaboratory & ScientificNew York State Department of Health
The full text, documents and contacts of this bid come with a ProposalWorkspace plan.Get started

At a glance

New York State Department of Health (New York) posted this RFP on Nov 6, 2026. Responses are due Jan 5, 2027. Category: Laboratory & Scientific. Questions are due 2026-11-24 16:30. Solicitation number 2140608.

Solicitation details

General information

Solicitation number
2140608
Notice type
RFP
Place of performance
New York
Questions due
2026-11-24 16:30

Agency and category

Agency
New York State Department of Health
Category
025Laboratory & Scientific

About this listing

ProposalWorkspace summary

This bid was found on the agency's procurement portal. The portal's own text is not republished here; the summary below was written by ProposalWorkspace's AI from the bid documents. The full posting is on the portal - subscribe to get the link.

Summary

AI-read from the bid documents

Summary fields on this page were extracted by AI from the bid documents. Always verify dates, amounts and requirements against the official solicitation.

What the bid is for

The Department receives the State Indoor Radon Grant from the U.S. Environmental Protection Agency, the terms of which require that radon detectors (detectors) be distributed to the general public at low cost. The contract resulting from this RFP will provide for the distribution of these devices directly to individuals within New York State and bulk shipments to various non-profit entities, the Department and/or county health departments, as specified by the Department. Individual users will return the detectors for analysis to the contractor’s laboratory facility via Business Reply Mail. Reports of analyses will be provided to the persons requesting the test with copies going to the Department’s Bureau of Environmental Radiation Protection (BERP).

Scope

It is the intent of the Department to obtain accurate and economical short-term (48 hour to 7 day) radon testing kits and laboratory analysis for New York State residents. The method for such measurements will be Activated Charcoal Adsorption Devices (AC) in conjunction with laboratory gamma scintillation counting. The Department has chosen this method based on a combination of cost efficiency and validation experience.

Orders will vary for this service, depending on public demand. The Department anticipates demand to be approximately 6,000 detectors per year, but this may vary due to factors outside the control of the Department. Approximately fifty percent (50%) of the detectors shipped yearly will be shipped in bulk (more than one unit) to various non-profit entities, the Department and/or Local Health Departments. [...] The Department will provide the Contractor with an order file containing the names and addresses of the person(s) requesting radon detectors and the number of detectors to ship on a weekly basis.

Radon Detector Hardware

  • The Contractor must provide Detectors that meet the following specifications:
  • Detectors shall be open-faced canisters or diffusion barrier canisters that can be easily distributed by mail.
  • Detectors shall maintain their physical integrity during mailing. Contractor will replace any damaged detectors at no additional cost to the state.
  • Detector measurements shall have a minimum detectable concentration of 0.5 picocuries per liter (pCi/L).
  • Detectors shall include instructions for detector placement, information recording, and mailing, in both English and Spanish.
  • The Contractor will ship detector hardware directly to the recipient within fourteen (14) calendar days of receiving the new order file from the Department.

Radon Detector Analysis and Reports

  • The Contractor must analyze detectors by gamma spectroscopy.
  • The Contractor must send written reports of detector analysis results to the persons who used the detectors within fourteen (14) calendar days of the Contractor’s receipt of detectors for analysis.
  • The Contractor must notify the Department by telephone or email emergency notification of measured radon concentrations greater than or equal to 100 pCi/L within two (2) business days of completion of the detector analysis.

Additional information

**Submission:** Proposals are due on or before **1/5/2027 at 4:30 p.m. ET**. Questions must be submitted via email to radon@health.ny.gov.

The Department intends to award one (1) contract from this procurement. The term of the Contract is expected to be for a period of five (5) years, subject to the availability of sufficient funding, successful Contractor performance, and approvals from the New York State Attorney General (AG) and the Office of the State Comptroller (OSC). Anticipated Contract Start Date: 10/1/2027.

**Minimum qualification (required):** Certification from the Department’s ELAP for the evaluation of activated charcoal detectors for detecting radon in air.

The Department establishes an overall goal of 30% for M/WBE participation. Prior to the start of work under the Contract, the Contractor shall procure insurance of the types and in the amounts set forth in Attachment 8, the New York State Department of Health Contract, Section IV. Contract Insurance Requirements.